Specifications

Ventilation modes

A/C-PC

Assist control (pressure control)

A/C-VC

Assist control (volume control)

CPAP

Continuous positive airway pressure

PSV

Pressure support ventilation

S/T

Spontaneous/timed ventilation

SIMV-PC

Synchronized intermittent mandatory ventilation (pressure control)

SIMV-VC

Synchronized intermittent mandatory ventilation (volume control)

MPV-PC

Mouthpiece ventilation (pressure control)

MPV-VC

Mouthpiece ventilation (volume control)

Physical

Weight

5.2 kg

Size

16.5 cm D x 28.6 cm W x 24.5 cm H 6.48” D x 11.25” W x 9.65” H

Screen dimensions

8”, 20.32 cm

Oxygen

Low flow

0 to 30 l/min; maximum 10 psi

High pressure

280 to 600 kPa (41 to 87 psi)

Measured and displayed patient parameters

Tidal volume (Vti or Vte)

0 to 2000 ml

Minute ventilation (MinVent)

0 to 30 l/min

Leak

0 to 200 l/min

Respiratory rate (RR)

0 to 90 BPM

Peak inspiratory flow (PIF)

0 to 200 l/min

Peak inspiratory pressure (PIP)

0 to 90 cmH2O

Mean airway pressure

0 to 90 cmH2O

Percentage spontaneous triggered breaths (%Spont Trig)

0 to 100%

I:E ratio

9.9:1 to 1:9.9

Dynamic compliance (Dyn C)

1 to 100 ml/cmH2O

Dynamic resistance (Dyn R)

5 to 200 cmH2O/l/sec

Dynamic plateau pressure (Dyn Pplat)

0 to 90 cmH2O

Auto-PEEP

0 to 20 cmH2O

FiO2 with FiO2 sensor

21% to 100%

SpO2 with pulse oximeter accessory

0 to 100%

Pulse rate with pulse oximeter accessory

18 to 321 beats per minute

EtCO2 with CO2 accessory

0 to 150 mmHg

Electrical

AC input voltage

100V - 240V, 50/60 Hz, 1.7 - 0.6A

DC input voltage

12/24V 6.5A

Internal and detachable Li-on batteries

15 hours’ nominal total run time per method in IEC 80601-2-72 (7.5 hours each battery)

Charge time for detachable and internal battery

from 0% to 80%: 2.5 hours; from 0% to 100%: 3.5 hours

Alarms

Inspiratory Pressure

PEEP+1 to 89 cmH2O

Tidal Volume

OFF, 10 - 2000 ml

Minute Ventilation

OFF, 0.2 - 30 L/min

Respiratory Rate

OFF, 1 - 90 BPM

Circuit Disconnection

OFF, 5 - 60 sec

Apnea Interval

5 to 60 sec

No trigger

OFF; 0.5-15 min (ONLY in MPV)

Controls

AVAPS with passive circuit

PSV, S/T, and A/C-PC modes only

Tidal volume

35 - 2000 ml on Dual Limb and Active Flow circuits, 50 - 2000 ml on passive and active PAP circuits

Breath rate

0 - 80 BPM

PEEP

0 - 35 cmH20 for active circuits 3 - 25 cmH20 for passive circuits

EPAP/CPAP

3 to 25 cmH20

IPAP

3 to 60 cmH20

Pressure support/pressure control

0 to 60 cmH20

Inspiratory time

0.3 - 5.0s

Rise time

0 to 6

Triggering and cycling

Off, AutoTrak, Sensitive, AutoTrak, and Flow Trigger

Flow trigger sensitivity

0.5 - 9 l/min

Flow cycle sensitivity

10% - 90% of peak flow

Flow pattern

Square, Ramp

FiO2

21% - 100%

Inspiratory time min/max

0.3 to 3.0 sec

Backup ventilation

ON-OFF